1910/1920 Basic Specifications
(See datasheet for full specifications)
Measured Parameters and Basic Measurement Accuracy:
| Parameter | Measurement Range | Fast | Medium | Slow |
| Ls,Lp | 0.001 nH to 99.999 H | +/-0.5% | +/-0.25% | +/-0.1% |
| Cs,Cp | 1 pF to 9.9999 F | +/-0.5% | +/-0.25% | +/-0.1% |
| DF | 0.00001 to 99.999 | +/-0.005 | +/-0.0025 | +/-0.001 |
| Q | 0.00000 to 9999.9 | +/-0.005 | +/-0.0025 | +/-0.001 |
| Y,Gp,Bp | 10 nS to 9999.9 S | +/-0.005 | +/-0.0025 | +/-0.001 |
| |Z|,Rs,Rp,Xs,ESR | 0.0001 mΩ to 99.999 MΩ | +/-0.5% | +/-0.25% | +/-0.1% |
| Phase Angle | -180.00 to +179.99 degrees | +/-1.8° | +/-0.9° | +/-0.18° |
| DCR | 0.1 mVto 100 KV | +/-0.5% | +/-0.25% | +/-0.1% |
| DUT AC Voltage | 20 mV to 1.0 V | +/-2% | +/-2% | +/-2% |
| DUT AC Current | 1 μA to 150 mA | +/-2% | +/-2% | +/-2% |
| DUT DC Voltage | 20 mV to 1.0 V | +/-2% | +/-2% | +/-2% |
| DUT DC Current | 1 μA to 150 mA | +/-2% | +/-2% | +/-2% |
*At optimum test signal levels, frequencies, DUT values and without calibration uncertainty. Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q) Factors, Impedance Z , Admittance Y , Phase Angle (0-), Equivalent Series Resistance (ESR), Conductance (Gp), Reactance (Xs), Susceptance (Bp) Note: s = series, p = parallel, ESR equivalent to Rs
Test Frequency Range: 20 Hz to 1 MHz, Continuous
Test Frequency Resolution: 1 Hz from 20 Hz to 1 kHz, 4 digits >1 kHz
Test Frequency Accuracy: +/- (0.01% +0.02 Hz)
Measurement Speed: Fast: 40 meas/sec Medium: 8 meas/sec Slow 1 meas/sec
Ranging: Automatic,Range Hold or user selectable
Trigger: internal (automatic), External(via RS-232,IEEE-488.2, Handler interfaces or Manual
Source Impedance: 5, 25, 50, 100 ohms selectable
AC Test Signal Voltage: 20 mV to 1.0 V (open circuit) in 5 mV steps
DC Bias Current Internal: 0.0 to 1.0 A in 1 mA steps External: up to 20.0 A
Standard Interfaces: IEEE-488.2 • RS-232 • Handler • Serial Printer
Measurement Delay: Programmable from 0 to 1000 ms in 1 ms steps
Averaging: Programmable from 1 to 1000
Median Value: Averaged over last three measurements
Built-in automatic adjustment procedure
Test Terminals: Front panel, four terminal (BNC) Optional Test Fixtures Available
Dimensions: 17x 5.25x 16in (432x 133x 406 mm)