The Trerice Connection Headis available
with both Type J and Type K Thermocouples,
as well as RTD sensors. The weatherproofed
head provides a conduit connection and is
available in cast aluminum (screw cover),
polypropylene (flip cover) and stainless
steel (screw cover). The stem is either
welded directly to the 1/2NPT threaded
connection, or is spring loaded to provide
maximum sensitivity. The spring loaded stem
must always be installed in a thermowell.Model Stem Style Stem Length Hot Junction Connection Head Material
TJD Type J T/C Z 316SS, 1/4O.D. 02 21/2" Stem U Ungrounded (T/C) S Spring Loaded, A Aluminum
TKDType K T/C 04 4" Stem D 3 Wire (RTD) 1/2NPT P Polypropylene
TDD100ΩRTD 06 6" Stem W Welded, S Stainless Steel
TMD1000ΩRTD 09 9" Stem 1/2NPT
450B、450LFB、450SS C、450LFSS D、450M E、450LFM 500XB、500XSS 700B、700LFB、700SS、700LFSS、700M、700LFM D82B、D82LFB、D83SS、D83LFSS 690B、691B、692B、693B、694B、696B600CB、610CB、620B、800LFB、760B、765B、766SS、700TA、SX9、HT30、HT31、AX1、BX1、CX1、AX2、BX2、CX2、BX9、4350、4352、B831X The Trerice Adjustable Angle
Bimetal Thermometercan be
configured to the most desirable
viewing angle. This instrument has
a hermetically sealed, stainless steel
case designed to withstand the
rigors of industrial environments,
while producing an accurate,
responsive measurement.
3", 5" Dial Size
± 1.0% Full Scale
Accuracy
Stainless Steel
Case & Stem
External Reset
Adjustable Angle
Models Dial Sizes
B836 3"
B856 5"
Case 300 stainless steel,
hermetically sealed
Stem 300 stainless steel
1/4"diameter
Coil Bimetallic, silicone dampened
on ranges to 300˚F (148˚C),
above 300˚F not dampened
Connection Adjustable angle, 1/2NPT
Window Double strength glass
Pointer Balanced, black finish
Dial Face Aluminum, white background with
black and blue graduations
and markings
External Reset Yes
Accuracy ±1.0 % Full Scale ASME B40.3
Grade A
Approximate Shipping Weight
B836: 1.1 lbs [0.5 kg]
B856: 1.5 lbs [0.68 kg]
Specifications
BIMETAL THERMOMETERS
B85606 shown
other lengths available: Specify in inches (72"maximum)
• Optional features available:
Please consult the Options &
Accessories Section for details.
Thermowell
• For corrosive or pressure
applications, use of a thermowell
is recommended to prevent
damage to the thermometer and
facilitate its removal from the process
(refer to pages 163-169).
For correct use and application
of all Bimetallic thermometers,
please refer to the Bimetallic
Actuated Thermometer Standard
ASME B40.3.