This is an IEC 60601 and IEC 80601 Defibrillation-proof tester. For the Energy Reduction test, joules consumed in the 100Ω resistor are immediately displayed after each pulse. The MegaPulse D8-PF tests protection against the effects of defibrillation (Common Mode, Differential Mode) and Energy reduction test (EC-13) and as noted in the Standards shown above. The MegaPulse D8-PF outputs a 360J Defib-proof test pulse per IEC 60601-1 for the Energy Reduction Test. It can also be used as a Defibrillator Lead Tester.
The most advanced medical surge tester available, the MegaPulse D8-PF is the latest iteration of the industry leading Defib-proof Tester from Compliance West. It adds the new Differential Mode Test shown in IEC 80601-2-26 and introduces our new touch controlled front panel to the legendary MegaPulse D5-PF defib proof tester. The MegaPulse D8-PF retains the direct energy readout and comprehensive list of included tests to the requirements of:
- IEC 80601-2-26:2019/COR1:2021 Figure 201.101 Defibrillation-Proof Applied Parts (S/N 435193 and up).
- IEC 60601-1:2005 (Issue 3) Figure 9 and 10 Defib-proof Differential- and Common-Mode Test (IEC 60601-1 Figure 50 in older versions).
- IEC 60601-1:2006 Figure 11 Energy Reduction Test.
- IEC 60601-2-4: 2010 Figure 201.110 Overload Test (optional).
- IEC 60601-2-34 Issue 2 (optional).
- IEC 60601-2-34 Issue 3 (included).
- IEC 60601-2-49.
- EC-13.
- EC-53.
- IEC 60601-2-25:2011.
- IEC 60601-2-27:2011.
-
Features
- Updated front panel with digital switches.
- Digital output voltage knob on the front panel.
- Integrated High Voltage on hold circuit.
- Integrated Energy measurement.
- Joules readout for the last pulse is displayed by pressing a button.
- Integrated Y1-Y2 measuring circuit.
- Built-in Pass/Fail reference circuit.
- External interlock disables HV output and defeats test when circuit is open.
- Long life capacitor rated for 2.5 million pulses.
- USB TestMinder, computer control.
- Auto-trigger via USB.
- Front panel indicates internal HV and polarity.
- Polarity button selector on the front panel.
- Front panel switches and instructions for various test setups.
- Microcontroller technologies.
- Leads, manual and calibration certificate included.
- One year calibration cycle.
-
Specifications
TESTING CIRCUITRY
Charge Voltage: 5100V ±1%.
Main Capacitance: 32uF ±5%.
Inductances: 500uH and 25mH ±5%.
Main Resistance (R1): 100Ω ±1% non-inductive.
Output Impedance: 50Ω and 400Ω ±1% non-inductive. Optional 10Ω available.
Voltage Control: Manually, by turning a knob in the front panel or via USB.
Polarity Control: Positive and Negative. Alternating control available with USB.
Voltage Display: 4 Digit LED Display.
Voltage Meter resolution: 5V.
Voltage Meter Accuracy: ±1%.
Duty Cycle: 20s between pulses, ±1% 100Ω resistor bank, up to 70 pulses.
12s between pulses, +1% <-5% 100Ω resistor bank, continuous duty -
ENERGY MEASUREMENT
Works only when: 25mH inductor selected (in accordance with the energy reduction test).
Display Resolution: 1 Joule.
Repeatability: ± 2 Joules between pulses.
Reading: From 3000V to 5100V only.ELECTRICAL
Line Voltage: 120V AC, 50/60Hz.
230V AC ±10%, 50 Hz ±1 Hz (CE for EU).SAFETY
Safety Interlock 2-pin terminal block on rear panel; shorted for operation.OUTPUTS – FRONT
50Ω, 400Ω, AUX, MEAS. POINT: High Voltage insulated rigid 4mm socket – Red.
RETURN: High Voltage insulated rigid 4mm socket – Black.
Y1, Y2, SINE WAVE: BNC.OUTPUTS – REAR
ENERGY MEASUREMENT: High Voltage insulated rigid 4mm socket – Red.
GROUND: High Voltage insulated rigid 4mm socket – Black.
TESTMNINDER: DB9 Female.
Inlet Module: Inlet module with fuse and power switch combination.